Abstract

We experimentally compare the performance and image contrast of spontaneous Raman and coherent Stokes Raman scattering microscopy. We demonstrate the differences between these techniques on a series of geometry-controlled samples that range in complexity from a point (array of tips) to one-dimensional (line grating) and, lastly, two-dimensional (checkereboard) microstructure. Through the use of this sample series, a comparison of the focal volume, achievable signal-to-noise, and resulting image contrast is made. The results demonstrate the effciency and spatial resolution attainable in coherent Raman microscopy relative to spontaneous Raman microscopy. Additionally, we detail potential complications in the interpretation of coherent Raman images of sample fine structure, where contrast is no longer based solely on oscillator concentration but can be influenced by sample microstructure. (C) 2011 Society of Photo-Optical Instrumentation Engineers (SPIE). [DOI:10.1117/1.3533310]

Publication Details
Publication Type
Journal Article
Year of Publication
2011
Volume
16
DOI
10.1117/1.3533310
Journal
Journal of Biomedical Optics